TECHNICAL CHARACTERISTICS MC 102

Trasducer input voltage: 5 Vdc 120 mA (max 8 cells 350 Ohm in parallel)
Measuring range: -3.9 ÷ +3.9 mV/V
Input sensitivity: 0.02 µV/ per count
Linearity: <0.01% of full scale
Gain drift: <0.0003% full scale/ °C
A/D Converter: 24 bit
Internal resolution: > 16.000.000 counts
Visible resolution: 99.999 division visible on net weight
Set decimal: 0.0 ; 0.00 ; 0.000; 0.0000
Filter: 0.2 ÷ 25 Hz adjustable
Tool voltage: 230 Vac ±10% - 50/60 Hz absorbed power 7 W (115 Vac on demand)
Temperature range: -10 ÷ +50°C
Storage temperature: -20 ÷ +70°C
Logic output: 4 relays (NA) Max. 115 Vac/30 Vdc 0.5 A each
Logic input: 3 optoisolated 12 Vdc/24 Vdc PNP
Serial door: COM1: RS232 half duplex COM2: RS422/RS485 half duplex
Transmission distance: 15m (RS232C), 1000m (RS422 e RS485)
Protocol: ASCII, Modbus, Printers Custom P190 o FT190, EPSON
Baud rate: 300 ÷ 115200 (adjustable)
Regulatory compliance: EN61000-6-3, EN1000-6-2, EN61010-1
Insulation: Class II
Dimensions: 144 x 72 x 120 mm (W x H x D)
Drilling template: 139 x 67 mm (W x H)
Electrical connection: Terminal screw pass 5.08 mm
Optional dosage software: 1 component in the loading or unloading up to 50 programmable recipes
Optional analog output: Optically isolated 16-Bit Voltage: 0÷5 or 0÷10V (R>10K Ohms); Current: 0÷20 o 4÷20mA (R < 300 Ohm); Linearity: 0.03% F.S.
BCD optional output: Throught external module
Profibus optional: Throught external module
Power supply: 12 Vdc; Analog output voltage available is 0÷5 V; Analog output current with R <260 Ohm; 24 VDC

Accessories are not to be considered included in the price of each product.